Services Available | |
---|---|
Repair | No |
Calibration | No |
Free Support | No |
The CS547 (pn 10865) measured conductivity by three passivated 316 stainless steel annular electrodes with dc isolation capacitors using the A547 and a data logger. A thermistor measured water temperature.
Read MoreThe "-L" on a product model indicates that the cable length is specified at the time of order.
The CS547 had a pH range of 3.0 to 9.0, with a conductivity accuracy of ±5% (at 25°C) over 0.45 to 7.0 mS/cm range, ±10% (at 25°C) over 0.005 to 0.45 mS/cm range. Temperature polynomial linearization accuracy was typically less than ±0.1°C over 0° to +48°C range. Interchangeability accuracy was typically less than ±0.2°C over 0° to +50°C range.
Operating Temperature | 0° to 50°C |
Weight | 0.3 kg (0.6 lb) with 3.05-m (10-ft) cable |
The CS547-L required one differential analog input, one single-ended analog input, and two excitation channels.