The AVW216 is a vibrating wire analyzer module that includes an internal 2.4 GHz spread-spectrum radio for wireless communication in many countries worldwide. With this vibrating wire analyzer module, your data logger can measure vibrating wire strain gages, pressure transducers, piezometers, tiltmeters, crackmeters, and load cells. These sensors are used in a wide variety of structural, hydrological, and geotechnical applications because of their stability, accuracy, and durability.
Read MorePurchase of this product is not recommended for new networks to be deployed in the European Union (EU) that may require future expansion. This and other RF-compatible products will not be available for sale in Europe after 01 Jan 2015 due to changes in EU legislation.
The AVW216 uses vibrating wire spectral-analysis technology (VSPECT®). VSPECT® observes the incoming sensor signal, performs a Fourier transform and a spectral analysis (transforming the time series into individual sinusoidal components in the frequency spectrum), and determines the sensor frequency by identifying the largest signal in the acceptable range while filtering out environmental and electrical noise.
The AVW216 analyzer module also provides many self-checking diagnostics such as vibrating element signal strength, signal-to-noise ratio, vibrating element signal decay ratio, and incorrect signal response. These diagnostics can be running in the background to give continual feedback of the condition for each sensor.
The AVW216 typically transmits its data to an RF416 spread-spectrum radio that is connected to the data logger. The AVW216 can also be connected directly to the data logger.
-NOTE- | Electrical specifications are valid over a -25° to +50°C range unless otherwise specified. Non-condensing environment required. |
Operating Temperature Range | -25° to +50°C |
Internal Radio Frequency Range | 2.450 to 2.482 GHz |
Radio Power | 50 mW |
Number of Vibrating Wire Sensors Measured | Up to 2 vibrating wire sensors can be connected to the analyzer module. Additional sensors can be measured by using an AM16/32-series multiplexer. |
Power Requirements | 9.6 to 16 Vdc |
Analog Input/Outputs | 2 differential (DF) vibrating wire measurements (V+ and V-) and 2 single-ended (SE) ratiometric resistive half-bridge measurements (T+ and T-) for vibrating wire sensor's onboard temperature sensor. |
Digital Control Ports |
3 digital control ports (C1 – C3)
|
RS-232 Port | 1 9-pin RS-232 port (for connecting to a data logger COM port) |
Measurement Resolution | 0.001 Hz RMS (±250 mV differential input range; -55° to +85°C) |
Measurement Accuracy | ±0.013% of reading (±250 mV differential input range; -55° to +85°C) |
Input Voltage Range | ±250 mV (differential) for vibrating wire inputs |
Common Mode Range | ±25 V |
Baud Rates | Selectable from 1200 to 38.4 kbps (ASCII protocol is one start bit, one stop bit, eight data bits, and no parity.) |
Memory |
|
Dimensions | 21.6 x 11.18 x 3.18 cm (8.5 x 4.5 x 1.2 in.) |
Weight | 0.43 kg (0.95 lb) |
Typical Current Drain @ 12 Vdc |
|
Quiescent, Radio Off | ~0.3 mA |
Radio Duty Cycling 1 s | ~3 mA (includes quiescent current) |
Radio Always On | ~26 mA (radio transmit current 100 mA) |
Active RS-232 Communication | ~6 mA (3 s after communication stops, the current will drop to the quiescent current) |
Measurement | ~25 mA (averaged over the 2 s) |
Note: The following shows notable compatibility information. It is not a comprehensive list of all compatible or incompatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | The CR10 supports the SDI-12 mode only. | |
CR1000 (retired) | ||
CR1000X | ||
CR10X (retired) | The CR10X supports the SDI-12 mode only. | |
CR200X (retired) | ||
CR206X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | The CR23X supports the SDI-12 mode only. | |
CR295X (retired) | ||
CR3000 (retired) | ||
CR500 (retired) | ||
CR5000 (retired) | The CR5000 supports the SDI-12 mode only. | |
CR510 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000C (retired) | ||
CR9000X (retired) |
Product | Compatible | Note |
---|---|---|
ENC10/12 | If using the RS-232 port, the AVW216 must be mounted flat to fit in an ENC10/12. | |
ENC10/12R | If using the RS-232 port, the AVW216 must be mounted flat to fit in an ENC10/12R. | |
ENC12/14 | If using the RS-232 port, the AVW216 must be mounted flat to fit in an ENC12/14. | |
ENC14/16 | ||
ENC16/18 | ||
ENC24/30 | ||
ENC24/30S |
The AVW216 Vibrating Wire Spectrum Analyzer Module allows our data loggers to measure vibrating wire strain gages, pressure transducers, piezometers, tiltmeters, crackmeters, and load cells.
The AVW216's internal spread-spectrum radio can transmit data to other AVW216 analyzer modules, RF416 spread-spectrum radios, and RF432 spread-spectrum radios. The AVW216 does not have a transparent mode, and therefore its internal radio cannot communicate with the RF415 radio.
The AVW216 requires a desiccated, non-condensing environment. A Campbell Scientific enclosure is recommended when the analyzer module is in the field. The AVW216 has built-in keyhole flanges for mounting to an enclosure backplate.
The AVW216 is typically powered by the data logger's power supply.
Current AVW200 firmware. Use the Device Configuration Utility version 1.13 or greater to send firmware and to configure the AVW200.
Number of FAQs related to AVW216: 12
Expand AllCollapse All
No. An AVW206, AVW211, or AVW216 is not capable of communicating with other AVW200-series modules.
The AVW200-series modules are not required, but they do reduce the chance of erroneous readings by filtering electrical noise that is sometimes present.
The AVW200 must be wired directly to the data logger via a cable.
The AVW206, AVW2011, and AVW216 have built-in spread-spectrum radios tuned to different frequencies that allow the data logger to communicate with a remote AVW device through an appropriate radio link.
It takes 2 s for the AVW200-series modules to acquire a measurement from each sensor. An AVW200-series module servicing two AM16/32B multiplexers, each with 16 vibrating wire sensors, requires approximately 64 s to acquire data.
Two probable causes are the following:
Use pn 17855, Data Cable, RS-232 Male to Pigtail.
Yes, using a null modem cable.
Two multiplexers, each with a maximum of 16 or 32 sensors per multiplexer, can be attached. If temperature measurements are excluded from each sensor, 32 sensors are possible.